FREESTYLE LITE
Report
- Report Number
- 2954323-2011-01527
- Event Type
- Malfunction
- Date Received
- February 8, 2011
- Date of Event
- January 13, 2011
- Report Date
- April 12, 2011
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
METER SERIAL NUMBER (B)(4) AND TEST STRIPS LOT NUMBER 1018911 HAVE BEEN RETURNED AND INVESTIGATED. THE COMPLAINT WAS NOT CONFIRMED. AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TESTS RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE FOUND IN METER MEMORY.
THIS IS A CORRECTION FOLLOW UP. FOLLOW UP NUMBER 1 IN HAD INCOMPLETE LOT NUMBER 101891. THE CORRECT LOT NUMBER IS 1018911.
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 2.7 MMOL/L, 6 MMOL/L, 1.1 MMOL/L, AND 3.7 MMOL/L WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 1018911 |