FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 1981861 · Received February 8, 2011

Report

Report Number
2954323-2011-01527
Event Type
Malfunction
Date Received
February 8, 2011
Date of Event
January 13, 2011
Report Date
April 12, 2011
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

METER SERIAL NUMBER (B)(4) AND TEST STRIPS LOT NUMBER 1018911 HAVE BEEN RETURNED AND INVESTIGATED. THE COMPLAINT WAS NOT CONFIRMED. AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TESTS RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE FOUND IN METER MEMORY.

Additional Manufacturer Narrative · 1

THIS IS A CORRECTION FOLLOW UP. FOLLOW UP NUMBER 1 IN HAD INCOMPLETE LOT NUMBER 101891. THE CORRECT LOT NUMBER IS 1018911.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 2.7 MMOL/L, 6 MMOL/L, 1.1 MMOL/L, AND 3.7 MMOL/L WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.