STARCLOSE SE VASCULAR CLOSURE SYSTEM
Report
- Report Number
- 2024168-2011-00758
- Event Type
- Injury
- Date Received
- February 8, 2011
- Date of Event
- November 8, 2010
- Report Date
- January 14, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE.
(B)(4). EVALUATION OF THE RETURNED DEVICE INDICATED IT WAS RETURNED FULLY CLIP-DEPLOYED. THE LOCATOR WINGS WERE FOUND BENT, WHICH LIKELY PREVENTED THEM FROM FULLY COLLAPSING INTO THE CLIP DELIVERY TUBESET DURING CLIP DEPLOYMENT. BECAUSE THE LOCATOR WINGS DID NOT COMPLETELY COLLAPSE INTO THE CLIP DELIVERY TUBESET, THIS DIRECTLY RESULTED IN THE DIFFICULTLY ENCOUNTERED REMOVING THE DEVICE. THE DEVICE WAS SUCCESSFULLY REMOVED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU) BY UTILIZING THE ACCESS PORTS TO RELEASE THE LOCKING MECHANISM ON THE THUMB ADVANCER. BASED ON THE INVESTIGATION FINDINGS, THE PROBABLE ROOT CAUSE FOR THE BENT LOCATOR WINGS THAT DIRECTLY RESULTED IN THE REPORTED DIFFICULTY ENCOUNTERED REMOVING THE DEVICE IS TISSUE COMPACTION THAT EXERTED A DISTAL FORCE ON THE LOCATOR WINGS WHILE IN THE TISSUE TRACT. SUBCUTANEOUS TISSUE TRAPPED BETWEEN THE DISTAL END OF THE TUBESET AND THE LOCATOR WINGS CAN BEND THE LOCATOR WINGS AND INTERFERE WITH CLIP DEPLOYMENT. NO MANUFACTURING OR QUALITY ISSUE WAS DETECTED. A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, AFTER CLIP DEPLOYMENT, DIFFICULTY WAS ENCOUNTERED REMOVING THE DEVICE. IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, A DILATOR WAS INSERTED INTO THE ACCESS PORTS UNLOCKING THE THUMB ADVANCER AND CLIP DELIVERY TUBESET ENABLING THEM TO BE RETRACTED PROXIMALLY, WHICH RELEASED THE DEVICE FROM THE TISSUE TRACT. WHEN THE DEVICE WAS REMOVED, BLEEDING CONTINUED. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STARCLOSE SE VASCULAR CLOSURE SYSTEM | IMPLANTABLE CLIP | MGB | AV-TEMECULA-CT | 930336H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |