ECHELON*FLEX60
Report
- Report Number
- 1527736-2011-00039
- Event Type
- Malfunction
- Date Received
- February 8, 2011
- Date of Event
- December 8, 2010
- Report Date
- December 9, 2010
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). CLOSURE TRIGGER TOP. THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH A ECR60G RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. HOWEVER, THE FIRING TRIGGER WAS NOT WORKING PROPERLY ON EACH STROKE AS THE PAWL WAS NOT ENGAGING WITH THE DRIVE BAR AT REPEATED ATTEMPTS. THE DEVICE WAS DISASSEMBLED TO VERIFY THE INTERNAL COMPONENTS AND WEAR WAS FOUND ON THE RIGHT SIDE OF THE CLOSURE TRIGGER DUE TO THE CLAMP FIRST LOCKOUT PIN ON THE GEARED TRIGGER PLATE. ONE POSSIBLE SCENARIO FOR THE DESCRIBED EVENT IS DUE TO APPLYING A LARGE PRYING FORCE ON THE CLOSURE TRIGGER HANDLE IN THE OPENING DIRECTION. THIS CAN THEN RESULT IN DAMAGE TO THE CLOSURE TRIGGER TOP COMPONENT IF THE APPLIED LOAD IS HIGH ENOUGH. ONCE THE CLOSURE TRIGGER TOP COMPONENT IS DAMAGED, THEN ANY ADDITIONAL ACTUATION OF THE FIRING TRIGGER CAN CAUSE IT TO RUB AGAINST THE NOW BROKEN OR BENT CLOSURE TRIGGER ASSEMBLY. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A THORACOTOMY PROCEDURE, THE DEVICE JAMMED AFTER FIVE FIRINGS. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED. NO DETAILS ARE AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON*FLEX60 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO SURGERY, INC (CINCINNATI) | UNK | G4RT8F |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |