FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 1981581 · Received January 31, 2011

Report

Report Number
9680959-2011-00204
Event Type
Malfunction
Date Received
January 31, 2011
Date of Event
November 12, 2010
Report Date
January 31, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE FRAME GRABBER BOARD. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INTERMITTENTLY THE IMAGE WILL DISAPPEAR. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1