LUCEA 50/100
Report
- Report Number
- 9710055-2024-00431
- Event Type
- Malfunction
- Date Received
- July 24, 2024
- Report Date
- July 24, 2024
- Manufacturer
- MAQUET SAS
- Product Code
- FTD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
INITIAL REPORTER WAS (B)(4). THE CORRECTION OF B5 DESCRIBE EVENT AND PROBLEM, H6 INVESTIGATION FINDINGS, H6 INVESTIGATION CONCLUSIONS DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS B5 DESCRIBE EVENT AND PROBLEM: ON 18TH JULY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 50 MOBILE. IT WAS STATED THE SPRING ARM COVER WAS MISSING. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY. CORRECTED B5 DESCRIBE EVENT AND PROBLEM: ON 18TH JULY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 50 MOBILE. IT WAS STATED THE SPRING ARM COVER WAS MISSING. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY. FURTHER INFORMATION PROVIDED BY GETINGE EMPLOYEE INDICATED THAT NOTHING FELL FROM DEVICE. THE COVER WAS LOST DURING MOVEMENT OF EQUIPMENT IN THE HOSPITAL AND WAS NOT USED WITHOUT COVER. BASED ON ADDITIONAL INPUT FROM GETINGE EMPLOYEE IT WAS POSSIBLE TO DETERMINE THAT THE ISSUE INVESTIGATED HEREIN IS NOT SAFETY AND RISK RELATED, AS THERE WAS NO INDICATION OF MISSING PARTS THAT COULD FALL OFF INTO STERILE FIELD, WHICH WAS INITIALLY CONSIDERED. THEREFORE, THE SCENARIO DESCRIBED IN THE RECORD IS CONSIDERED AS NON-REPORTABLE. PREVIOUS H6 INVESTIGATION FINDINGS: RESULTS PENDING COMPLETION OF INVESTIGATION///3233. CORRECTED H6 INVESTIGATION FINDINGS: NONE. PREVIOUS H6 INVESTIGATION CONCLUSIONS: CONCLUSION NOT YET AVAILABLE//11. CORRECTED H6 INVESTIGATION CONCLUSIONS: CAUSE TRACED TO USER//19. INITIALLY PROVIDED INFORMATION WAS POINTING TO MISSING SPRING ARM COVER. THE ISSUE IS CONSIDERED AS SAFETY RELATED AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY. ACCORDING TO ADDITIONAL CLARIFICATION PROVIDED BY THE GETINGE TECHNICIAN, THE INITIAL INFORMATION WAS INCORRECT. IT WAS DETERMINED THAT THE ISSUE INVESTIGATED HEREIN IS NOT SAFETY AND RISK RELATED AS THERE WAS NO INDICATION OF MISSING PARTS THAT COULD FALL OFF INTO STERILE FIELD. THE INVESTIGATION WAS PERFORMED. THE INVESTIGATED SCENARIOS DID NOT CAUSE RISK TO HUMAN LIFE. THE REVIEW OF THE CUSTOMER PRODUCT COMPLAINTS, RELATED TO INVESTIGATED ISSUE IN TIME, SHOWS THAT THERE IS NO REGULAR INCOME. IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE DEVICE WAS NOT BEING USED FOR PATIENT TREATMENT OR DIAGNOSIS. NO APPARENT REASON WAS IDENTIFIED FOR SUGGESTING TO OPEN A CAPA OR EVALUATION FOR THE NEED OF ANOTHER ACTION IN THE MARKET.
EVENT SITE NAME: 200151-08 - (B)(6). ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION.
ON 18TH JULY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 50 MOBILE. IT WAS STATED THE SPRING ARM COVER WAS MISSING. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIUS INJURY. FURTHER INFORMATION PROVIDED BY GETINGE EMPLOYEE INDICATED THAT NOTHING FELL FROM DEVICE. THE COVER WAS LOST DURING MOVEMENT OF EQUIPMENT IN THE HOSPITAL AND WAS NOT USED WITHOUT COVER. BASED ON ADDITIONAL INPUT FROM GETINGE EMPLOYEE IT WAS POSSIBLE TO DETERMINE THAT THE ISSUE INVESTIGATED HEREIN IS NOT SAFETY AND RISK RELATED, AS THERE WAS NO INDICATION OF MISSING PARTS THAT COULD FALL OFF INTO STERILE FIELD, WHICH WAS INITIALLY CONSIDERED. THEREFORE, THE SCENARIO DESCRIBED IN THE RECORD IS CONSIDERED AS NON-REPORTABLE.
ON 18TH JULY, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 50 MOBILE. IT WAS STATED THE SPRING ARM COVER WAS MISSING. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 364572 | LUCEA 50/100 | LAMP, SURGICAL | FTD | MAQUET SAS | ARDLCA309005A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |