FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1981394 · Received December 30, 2010

Report

Report Number
2531779-2010-03231
Event Type
Malfunction
Date Received
December 30, 2010
Report Date
December 3, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL OF THE DEVICE CONFIRMED THE ALLEGED ISSUE. THE DOWNLOAD HISTORY INDICATED THAT THE PT PERFORMED A BATTERY CHANGE AFTER A "LOW BATTERY" ALARM OCCURRED ON (B)(6) 2010 AT 21:04. THE BATTERY CHANGE RESULTED WITH THE PUMP RESETTING TO A DEFAULT DATE/TIME OF (B)(6) 2007 AND 12:00 AM. THE PUMP WAS UNABLE TO HOLD THE DATE/TIME AND RESET ITSELF TO THE DEFAULT DATE/TIME. THE PUMP WAS OPENED FOR FURTHER ANALYSIS AND A LEAKING BT1 COMPONENT WAS OBSERVED.

Description of Event or Problem · 1

ON (B)(6) 2010, THE PT CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS RESETTING TO THE DEFAULT DATE/TIME ON IT'S OWN WITHOUT BATTERY REMOVAL. THE PT DENIED HAVING ANY BLOOD GLUCOSE EXCURSIONS DUE TO THE ISSUE. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGATION THAT THE PUMP WAS RESETTING ON IT'S OWN TO THE DEFAULT DATE/TIME. THERE IS NO EVIDENCE; HOWEVER, THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PT DID NOT ALLEGE ANY HARM OR INJURY DUE TO THE ALLEGED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1