FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1981329 · Received February 2, 2011

Report

Report Number
1824206-2011-00617
Event Type
Malfunction
Date Received
February 2, 2011
Date of Event
January 19, 2011
Report Date
January 19, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THAT THE SIDE COMMUNICATION CABLE WAS NOT WORKING. FOUND CABLE WAS TORN AND FUSE F9 WAS NO GOOD. SIDE COMMUNICATION CABLE WAS TORN AND COPPER WIRES EXPOSED. REPLACED SIDE COMMUNICATION CABLE AND F9 FUSE ON PCB BOARD TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

ACCOUNT ALLEGES THAT THE TOTALCARE SIDE COMM CABLE WAS NOT WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1900C003210

Patients

Seq Age Sex Outcome Treatment
1