FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 1981329
·
Received February 2, 2011
Report
- Report Number
- 1824206-2011-00617
- Event Type
- Malfunction
- Date Received
- February 2, 2011
- Date of Event
- January 19, 2011
- Report Date
- January 19, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND THAT THE SIDE COMMUNICATION CABLE WAS NOT WORKING. FOUND CABLE WAS TORN AND FUSE F9 WAS NO GOOD. SIDE COMMUNICATION CABLE WAS TORN AND COPPER WIRES EXPOSED. REPLACED SIDE COMMUNICATION CABLE AND F9 FUSE ON PCB BOARD TO RESOLVE THIS ISSUE.
Description of Event or Problem · 1
ACCOUNT ALLEGES THAT THE TOTALCARE SIDE COMM CABLE WAS NOT WORKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | P1900C003210 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |