FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1981302 · Received February 2, 2011

Report

Report Number
1824206-2011-00616
Event Type
Malfunction
Date Received
February 2, 2011
Date of Event
January 13, 2011
Report Date
January 13, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THE HEAD OF BED NOT GOING UP EVENLY. REPLACED THE PIVOT SLIDE AND THE SLIDER EXTRUSION TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

COMPLAINT ALLEGED THAT THE HEAD SECTION OF THIS BED WAS BENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1840BCA000025

Patients

Seq Age Sex Outcome Treatment
1