FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE WITH INTEGRATED HOLDER

MDR report key: 19812258 · Received July 23, 2024

Report

Report Number
9617032-2024-01110
Event Type
Malfunction
Date Received
July 23, 2024
Date of Event
June 10, 2024
Report Date
July 10, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
FMI
UDI-DI
30382903688358
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: "MATERIAL #: 368835. LOT/BATCH #: 4030223. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL DRAW TESTING AND NO ISSUES WERE OBSERVED RELATING TO HUB-HOLDER SEPARATION AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE HUB-HOLDER SEPARATION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.".

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING A BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE WITH INTEGRATED HOLDER, THE TUBE HOLDER DETACHED FROM THE REST OF THE DEVICE WHEN A BLOOD COLLECTION TUBE WAS INSERTED. NO IMPACT WAS REPORTED. REPORT 5 OF 5.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1004623 BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE WITH INTEGRATED HOLDER BLOOD SPECIMEN COLLECTION DEVICE FMI BECTON, DICKINSON AND COMPANY (BD) 4030223 30382903688358

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown