FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 CXE VOLUMETRICINFUSION PUMP

MDR report key: 1981135 · Received February 8, 2011

Report

Report Number
6000001-2011-00731
Event Type
Malfunction
Date Received
February 8, 2011
Date of Event
October 13, 2010
Report Date
October 15, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE CONDITION OF BATTERY DEPLETED SET ALARM WAS CONFIRMED THROUGH THE EVENT HISTORY. THE ROOT CAUSE OF THE REPORTED PROBLEM COULD NOT BE IDENTIFIED; HOWEVER THE MAIN BATTERIES AND HARNESS WERE REPLACED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). (B)(4).

Description of Event or Problem · 1

THIS IS A REPORT FROM BAXTER (B)(4) OF A BATTERY FAILURE ALARM. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. IT IS UNKNOWN IF PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER CANADA PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED SET WHICH INTERRUPTED DELIVERY. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.92 CATEGORIZED AS REMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3 CXE VOLUMETRICINFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1