COLLEAGUE 3 CXE VOLUMETRICINFUSION PUMP
Report
- Report Number
- 6000001-2011-00731
- Event Type
- Malfunction
- Date Received
- February 8, 2011
- Date of Event
- October 13, 2010
- Report Date
- October 15, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE CONDITION OF BATTERY DEPLETED SET ALARM WAS CONFIRMED THROUGH THE EVENT HISTORY. THE ROOT CAUSE OF THE REPORTED PROBLEM COULD NOT BE IDENTIFIED; HOWEVER THE MAIN BATTERIES AND HARNESS WERE REPLACED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).
ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). (B)(4).
THIS IS A REPORT FROM BAXTER (B)(4) OF A BATTERY FAILURE ALARM. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. IT IS UNKNOWN IF PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER CANADA PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED SET WHICH INTERRUPTED DELIVERY. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.92 CATEGORIZED AS REMEDIATED.