FDA Adverse Event Injury Summary report: N

COGNIS

MDR report key: 1980976 · Received February 8, 2011

Report

Report Number
2124215-2011-02162
Event Type
Injury
Date Received
February 8, 2011
Date of Event
November 29, 2010
Report Date
November 29, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE PHYSICIAN ELECTED TO REMOVE THE RV LEAD FROM SERVICE ONE DAY POST-IMPLANT. THE REST OF THE SYSTEM WAS ALSO REMOVED AS A RESULT, FOR IN ATTEMPTING TO REMOVE THIS RV LEAD, BOTH THE NON-BOSTON SCIENTIFIC RIGHT ATRIAL (RA) LEAD AND THE LEFT VENTRICULAR (LV) LEADS BECAME DISLODGED. IT WAS OBSERVED THAT THE LV LEAD THEN BEGAN TO UNRAVEL, LEADING TO COMPLICATION WITH REMOVING THE LV LEAD, RESULTING IN RESULT DISSECTION AND OCCLUSION OF THE SUPERIOR VENA CAVA. THE LV LEAD WAS ULTIMATELY REMOVED IN PIECES. THE RA LEAD WAS REMOVED WITHOUT ISSUE. THE ASSOCIATED CRT-D WAS ALSO REMOVED WITHOUT ISSUE, BEYOND THE ALLEGATION OF RV UNDERSENSING AND ATTEMPTED RE-PROGRAMMING. THE PATIENT REMAINED HOSPITALIZED FOR MONITORING, AND WAS NOTED AS STABLE. MOST RECENTLY, THE PATIENT HAD NO SYSTEM IMPLANTED. A LEFT-SIDED IMPLANT WAS BEING EVALUATED. THIS DEVICE HAS BEEN RETURNED. BOSTON SCIENTIFIC HAS CONCLUDED THAT NO DEVICE CHARACTERISTICS WERE IDENTIFIED THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, INVESTIGATION IS CONSIDERED COMPLETE. THIS REPORT WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) IN ASSOCIATION WITH THE TRANSVENOUS RIGHT VENTRICULAR (RV) LEAD WAS REMOVED FROM SERVICE ALONG WITH THE REST OF THE PATIENT'S DEVICE SYSTEM ONE DAY POST-IMPLANT. ONLY TWO OF THE PRODUCTS IN THE DEVICE SYSTEM WERE INVOLVED FOR POTENTIALLY LIFE THREATENING OCCURRENCE, AS FOLLOWS: DURING IMPLANT OF THE ACUTE DEVICE SYSTEM, THIS RV LEAD IN ASSOCIATION WITH THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DID EXHIBIT RV UNDERSENSING. THIS UNDERSENSING WAS THE RESULT OF ATRIAL OVERSENSING OF THE RV WHICH IN TURN LED TO BLANKING OF THE RV. THE UNDERSENSING DID DELAY THERAPY FOR AN UNKNOWN DURATION. THE PHYSICIAN ATTEMPTED TO CORRECT THIS ISSUE WITH RE-PROGRAMMING DURING THE SAME PROCEDURE. IT REMAINS UNCLEAR WHETHER THERE WAS MORE UNDERSENSING POST-ADJUSTMENT OF THE BLANKING PERIOD. THE PATIENT WAS NOTED TO HAVE A HISTORY OF HIGH DEFIBRILLATION THRESHOLD (DFTS) TESTING MEASUREMENTS ALSO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND N118

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention N118| 4518| 0184| 5076| H219