UNICEL® DXI 800 ACCESS IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-00335
- Event Type
- Malfunction
- Date Received
- February 7, 2011
- Date of Event
- October 14, 2010
- Report Date
- October 14, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JLB
- PMA / PMN Number
- K023764
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
THE SAMPLES WERE COLLECTED IN LI HEP PLASMA TUBES WITHOUT GEL SEPARATOR. PER CUSTOMER, THE SAMPLES WERE FULL DRAWS AND DID NOT APPEAR LIPEMIC. ALL THREE LEVELS OF QC WERE WITHIN THE CUSTOMER'S ESTABLISHED RANGES. A BCI FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED THE SUBSTRATE PROBE AND TUBING. FSE ALSO ADJUSTED THE BUBBLE DETECTOR AND PERFORMED HIGH SENSITIVITY SYSTEM CHECK, WHICH PASSED WITHIN INSTRUMENT SPECIFICATIONS. THE FSE RECALIBRATED THE ASSAY AND PERFORMED QC; NO ISSUES WERE REPORTED. A CLEAR ROOT CAUSE CAN NOT BE DETERMINED FOR THIS EVENT. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS WITHIN THE DATE RANGE (B)(4) 2010 - (B)(4) 2010 FOR ADDITIONAL REPORTABLE EVENTS/OCCURRENCES. THIS REPORTABLE EVENT CAPTURES PATIENT #2 RESULTS THAT ARE RELATED TO THE PREVIOUSLY SUBMITTED MDR 2122870-2010-00745.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO OBTAINING AN ERRONEOUSLY ELEVATED CREATININE KINASE - MB (CKMB) RESULT ABOVE THE NORMAL REFERENCE RANGE FOR ONE PATIENT THAT WAS GENERATED BY THE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. SUBSEQUENT TESTING PRODUCED LOWER RESULTS WITHIN THE NORMAL REFERENCE RANGE. THE ERRONEOUS RESULT WAS REPORTED OUT OF THE LABORATORY. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OF CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXI 800 ACCESS IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JLB | BECKMAN COULTER INC. | DXI 800 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |