FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 1980513
·
Received February 7, 2011
Report
- Report Number
- 2050012-2011-00385
- Event Type
- Malfunction
- Date Received
- February 7, 2011
- Date of Event
- January 18, 2011
- Report Date
- January 18, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) REPLACED SYRINGES THAT APPEARED TO BE LEAKING PAST THE PLUNGER AND HAVE BUFFER BUILD UP. FSE REROUTED THE OVERFLOW TRAY TUBING, WHICH HAD CAUSED FLUID IN THE TRAY TO NOT DRAIN PROPERLY. REPLACED LEAKING WASH STATION RINSE AND DRAIN TUBING.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO CLOSE TUBE ACCESSION (CTA) WASH STATION LEAK. NO AFFECT TO ANY PERSONNEL OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600I | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |