FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 1980513 · Received February 7, 2011

Report

Report Number
2050012-2011-00385
Event Type
Malfunction
Date Received
February 7, 2011
Date of Event
January 18, 2011
Report Date
January 18, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BCI FIELD SERVICE ENGINEER (FSE) REPLACED SYRINGES THAT APPEARED TO BE LEAKING PAST THE PLUNGER AND HAVE BUFFER BUILD UP. FSE REROUTED THE OVERFLOW TRAY TUBING, WHICH HAD CAUSED FLUID IN THE TRAY TO NOT DRAIN PROPERLY. REPLACED LEAKING WASH STATION RINSE AND DRAIN TUBING.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO CLOSE TUBE ACCESSION (CTA) WASH STATION LEAK. NO AFFECT TO ANY PERSONNEL OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600I N/A