OSS SEGMENTAL FEMORAL
Report
- Report Number
- 0001825034-2024-01886
- Event Type
- Injury
- Date Received
- July 23, 2024
- Date of Event
- March 5, 2024
- Report Date
- July 23, 2024
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- KRO
- UDI-DI
- 00880304238824
- PMA / PMN Number
- K052685
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). D10 - MEDICAL PRODUCT: OSS POROUS IM STEM CATALOG # 150381 LOT # 414120; OSS 4CM DIAPHYSEAL SEGMENT CATALOG # 150482 LOT # 899890; OSS POR STRAIGHT STEM 16.5X150 CATALOG # 150451 LOT # 183380; OSS NON-MOD TIB PLATE LONG 63 CATALOG # 150419 LOT # 443540; OSS TIBIAL POLY BEARING 14MM CATALOG # 150411 LOT # 063740; ALLEN MEDULLARY CEMENT PLUGS 1-16 MM DIAMETER FLANGE/8 MM DIAMETER CORE STORE IN COOL DRY PLACE CATALOG # 00801102016 LOT # 64325372. G2: AUSTRALIA. H6: SUGGESTED COMP CODE: MECHANICAL (04) -FEMUR. REPORTED EVENT WAS CONFIRMED BY REVIEW OF MEDICAL RECORDS PROVIDED. VISUAL EXAMINATION OF THE RETURNED PRODUCT FOUND THAT IT SHOWS SIGNS OF BEING IMPLANTED WEAR/DISCOLORATION / FOREIGN MATERIAL / SCRATCHED AND EXPLANTED. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. X-RAY REVIEW INDICATES THERE IS EXTENSIVE RADIOLUCENCY ALONG THE FEMORAL IMPLANT WHICH IS LOOSE AS NOTED. OVERALL ALIGNMENT OF THE RIGHT KNEE ARTHROPLASTY IS MAINTAINED IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE ONE YEAR POST IMPLANTATION DUE TO LOOSENING OF IMPLANT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 935869 | OSS SEGMENTAL FEMORAL | PROSTHESIS KNEE | KRO | ZIMMER BIOMET, INC. | N/A | 542380 | 00880304238824 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | Hospitalization| R | SEE H10 NARRATIVE. |