FDA Adverse Event Injury Summary report: N

OSS DIAPHYSEAL SEGMENT

MDR report key: 19805057 · Received July 23, 2024

Report

Report Number
0001825034-2024-01887
Event Type
Injury
Date Received
July 23, 2024
Date of Event
March 5, 2024
Report Date
July 23, 2024
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JDI
UDI-DI
00880304239982
PMA / PMN Number
K002757
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) . D10 - MEDICAL PRODUCT: OSS POROUS IM STEM CATALOG # 150381 LOT # 414120 OSS 7CM SEGMENTAL FEMORAL RT CATALOG # 150354 LOT # 542380 OSS POR STRAIGHT STEM 16.5X150 CATALOG # 150451 LOT # 183380 OSS NON-MOD TIB PLATE LONG 63 CATALOG # 150419 LOT # 443540 OSS TIBIAL POLY BEARING 14MM CATALOG # 150411 LOT # 063740 ALLEN MEDULLARY CEMENT PLUGS 1-16 MM DIAMETER FLANGE/8 MM DIAMETER CORE STORE IN COOL DRY PLACE CATALOG # 00801102016 LOT # 64325372 G2: AUSTRALIA REPORTED EVENT WAS CONFIRMED BY REVIEW OF MEDICAL RECORDS PROVIDED. VISUAL EXAMINATION OF THE RETURNED PRODUCT FOUND THAT IT SHOWS SIGNS OF BEING IMPLANTED DISCOLORATION / FOREIGN MATERIAL / SCRATCHED AND EXPLANTED. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. X-RAY REVIEW INDICATES THERE IS EXTENSIVE RADIOLUCENCY ALONG THE FEMORAL IMPLANT WHICH IS LOOSE AS NOTED. OVERALL ALIGNMENT OF THE RIGHT KNEE ARTHROPLASTY IS MAINTAINED IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE ONE YEAR POST IMPLANTATION DUE TO LOOSENING OF IMPLANT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
935861 OSS DIAPHYSEAL SEGMENT PROSTHESIS KNEE JDI ZIMMER BIOMET, INC. N/A 899890 00880304239982

Patients

Seq Age Sex Outcome Treatment
1 NA Female Hospitalization| R SEE H10 NARRATIVE