FDA Adverse Event Malfunction Summary report: N

PRECISION XTRA

MDR report key: 1980445 · Received February 7, 2011

Report

Report Number
2954323-2011-01498
Event Type
Malfunction
Date Received
February 7, 2011
Date of Event
January 13, 2011
Report Date
February 7, 2011
Product Code
NBW
Removal / Correction Number
ADC FA1197-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30C DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE C, D OR E ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. THE PRIMARY CAUSE HAS BEEN IDENTIFIED TO BE BATCH TO BATCH GLUE VARIABILITY. RESULTS EVALUATION: THE FDA HAS BEEN INFORMED OF THE FIELD ACTION PER 21CFR806 (FIELD ACTION REFERENCE NUMBER 2954323-12/22/10-001-R). ALL AFFECTED CONSIGNEES WERE NOTIFIED BY LETTER BEGINNING DECEMBER 22, 2010.

Description of Event or Problem · 1

IN USING AN AFFECTED TEST STRIP RELATED TO AN ON-GOING FIELD ACTION FOR ABBOTT PRECISION FAMILY TEST STRIPS, THE CUSTOMER REPORTED A READINGS ISSUE. AS A RESULT OF THIS ISSUE, THE CUSTOMER REPORTED FEELING DIZZY, EXPERIENCING A HEADACHE, AN INCREASED HEARTBEAT, BLURRED VISION, AND AN INCREASED URGENCY TO USE THE BATHROOM. THE CUSTOMER TOOK THEIR PRESCRIBED DOSAGE OF GLIPIZIDE IN ORDER TO COUNTERACT SYMPTOMS. THERE WAS NO REPORT OF THIRD PARTY MEDICAL INTERVENTION, NO MEDICATION ADMINISTERED OUTSIDE OF THEIR NORMAL TREATMENT REGIMEN, NOR WAS THERE DIAGNOSIS OF HYPOGLYCEMIA, HYPERGLYCEMIA, OR ANY OTHER DIABETES RELATED CONDITION. THERE WAS NO REPORT OF DEATH OR SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISION XTRA BLOOD GLUCOSE MONITORING SYSTEM NBW 45707

Patients

Seq Age Sex Outcome Treatment
1