FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19803310 · Received July 23, 2024

Report

Report Number
9610877-2024-55601
Event Type
Malfunction
Date Received
July 23, 2024
Date of Event
July 16, 2024
Report Date
July 23, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
PEA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EPK-3000-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K172156. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE SCOPE CONNECTOR MECHANICAL UNIT VIDEO IMAGE FREEZING. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE SCOPE CONNECTOR MECHANICAL UNIT. THIS DEFECT OCCURRED NOT DURING PROCEDURE. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2479295 PENTAX VIDEO PROCESSOR (INTERNATIONAL) PEA HOYA CORPORATION PENTAX TOKYO OFFICE EPK-3000

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown