FDA Adverse Event Injury Summary report: N

COMPRESSION ANASTOMOSIS RING (CAR 27)

MDR report key: 1980228 · Received February 1, 2011

Report

Report Number
3005278776-2011-00001
Event Type
Injury
Date Received
February 1, 2011
Date of Event
December 31, 2010
Report Date
January 2, 2011
Manufacturer
NITI SURGICAL SOLUTIONS LTD.
Product Code
FZP
PMA / PMN Number
K062008
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS SUBMITTED ON BEHALF OF THE MANUFACTURER (NITI SURGICAL SOLUTIONS LTD; (B)(4)) AND THE IMPORTER ((B)(4)), AS NITI SURGICAL SOLUTIONS LTD. SERVES AS THE DESIGNATED COMPLAINT HANDLING UNIT FOR BOTH FACILITIES. THE PRODUCTION HISTORY FILES WERE REVIEWED, INDICATING THAT THE DEVICE WAS RELEASED ACCORDING TO THE PRODUCT SPECIFICATIONS. THE RING WAS RETURNED AND INVESTIGATED. NO FAILURE WAS DETECTED AND THE RING WAS FOUND TO BE WITHIN ITS SPECIFICATION. ANASTOMOTIC LEAKAGE IS AN ANTICIPATED COMPLICATION OF COLORECTAL ANASTOMOTIC PROCEDURES AND THE CURRENT CUMULATIVE LEAK RATE FOUND WITH THE USE OF THE COLONRING DEVICE IS WITHIN THE RANGE REPORTED IN THE LITERATURE FOR ANASTOMOSIS DEVICES.

Description of Event or Problem · 1

PATIENT SUFFERING OF COLON MALIGNANCY UNDERWENT LAPAROSCOPIC LAR, ANASTOMOSIS WAS PERFORMED USING THE COLONRING DEVICE. ON POD 3, PATIENT HAD HIGH FEVER AND TENDER ABDOMEN. CT SCAN REVEALED ANASTOMOTIC LEAKAGE. RING WAS REMOVED AND HARTMAN PROCEDURE WAS CONDUCTED FOLLOWED BY ANTIBIOTIC TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPRESSION ANASTOMOSIS RING (CAR 27) IMPLANTABLE CLIP (FZP) FZP NITI SURGICAL SOLUTIONS LTD. CAR 27 43001006

Patients

Seq Age Sex Outcome Treatment
1 48 YR Required Intervention ALPRAZOLAM| ESTRADIOL| THYROXINE