FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE

MDR report key: 19801711 · Received July 22, 2024

Report

Report Number
9610773-2024-31791
Event Type
Malfunction
Date Received
July 22, 2024
Date of Event
June 21, 2024
Report Date
January 2, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761075015
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. UPDATED FIELDS: D8, H2, H3, H4, H6, H11. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: -INTERNAL CORROSION IN INSERTION TUBE. -INSERTION TUBE IS DENT. -INSERTION TUBE IS BEND. -LIGHT GUIDE BUNDLE IS BROKEN. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: CAUSED BY A COMPONENT FAILURE OF VIDEO CONNECTOR COVER. THE CAUSE OF VIDEO CONNECTOR COVER FAILURE COULD NOT BE DETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER. THE INTENDED PROCEDURE WAS COMPLETED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING AN UNSPECIFIED PROCEDURE THE SUBJECT DEVICE PRESENTED A DAMAGE ON ITS UNIVERSAL CONNECTING SECTION. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
935647 VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE RIGID VIDEO LAPAROSCOPE HET OLYMPUS WINTER & IBE GMBH WA50052A 04042761075015

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown