FDA Adverse Event Malfunction Summary report: N

TRI-STAPLE 2.0

MDR report key: 19801608 · Received July 22, 2024

Report

Report Number
1219930-2024-03302
Event Type
Malfunction
Date Received
July 22, 2024
Date of Event
June 25, 2024
Report Date
September 18, 2024
Manufacturer
COVIDIEN LP LLC NORTH HAVEN
Product Code
GDW
UDI-DI
10884521717626
PMA / PMN Number
K173270
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

D10 CONCOMITANT PRODUCT: SIGPHANDLE, SIG POWER SIGPHANDLE HANDLE (LOT#: UNKNOWN); SIGPSHELL, SIG POWER SIGPSHELL CONTROL SHELL (LOT#: UNKNOWN); SIGADAPTSTND, SIG POWER SIGADAPTSTND LINEAR ADAPTER (LOT#: UNKNOWN). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THAT THE RELOAD WAS PARTIALLY FIRED AND THE INTERLOCK WAS ENGAGED. FUNCTIONAL TESTING FOUND THAT THE RELOAD WAS LOADED INTO A HANDLE, AND THE CLAMPING MECHANISM WAS DEFORMED. IT WAS REPORTED THAT DURING A LAPAROSCOPIC DISTAL GASTRECTOMY (LDG), THE FIRST GASTRECTOMY FIRING FROM THE GREATER CURVATURE SIDE, THE DEVICE STOPPED AROUND MIDWAY. THE USER REMOVED THE MIDDLE OF THE REINFORCE SHEET AND REMOVED THE CARTRIDGE FROM THE STOMACH. THE REPORTED ISSUE WAS CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: FAILURE TO COMPLETELY FIRE THE RELOAD WILL RESULT IN AN INCOMPLETE CUT AND/OR INCOMPLETE STAPLE FORMATION, WHICH MAY RESULT IN POOR HEMOSTASIS AND/OR LEAKAGE. ALWAYS INSPECT THE TISSUE THICKNESS AND SELECT AN APPROPRIATE STAPLE SIZE PRIOR TO APPLICATION OF THE ENDO GIA ULTRA UNIVERSAL SHORT, ENDO GIA ULTRA UNIVERSAL OR ENDO GIA ULTRA UNIVERSAL XL STAPLER. OVERLY THICK OR THIN TISSUE MAY RESULT IN UNACCEPTABLE STAPLE FORMATION. WHEN POSITIONING THE STAPLER ON THE APPLICATION SITE, ENSURE THAT NO OBSTRUCTIONS, SUCH AS CLIPS, ARE INCORPORATED INTO THE INSTRUMENT JAWS. FIRING OVER AN OBSTRUCTION MAY RESULT IN INCOMPLETE CUTTING ACTION AND/OR IMPROPERLY FORMED STAPLES. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC DISTAL GASTRECTOMY (LDG), THE FIRST GASTRECTOMY FIRING FROM THE GREATER CURVATURE SIDE, THE DEVICE STOPPED AROUND MIDWAY. THE USER REMOVED THE MIDDLE OF THE REINFORCE SHEET AND REMOVED THE CARTRIDGE FROM THE STOMACH. A NEW RELOAD WAS USED TO RESOLVED THE ISSUE. NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2482637 TRI-STAPLE 2.0 STAPLE, IMPLANTABLE GDW COVIDIEN LP LLC NORTH HAVEN SIGTRSB60AMT N4C1866Y 10884521717626

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown