FDA Adverse Event
Malfunction
Summary report: N
IMMAGE® IMMUNOCHEMISTRY SYSTEM
MDR report key: 1979669
·
Received February 6, 2011
Report
- Report Number
- 2050012-2011-00353
- Event Type
- Malfunction
- Date Received
- February 6, 2011
- Date of Event
- January 12, 2011
- Report Date
- January 12, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JQX
- PMA / PMN Number
- K962294
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED: THE FSE INSPECTED THE INSTRUMENT AND FOUND OUT THAT WASH DRAIN TUBING CAME OFF AND RECONNECTED THE TUBING AND THIS ISSUE HAS BEEN RESOLVED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI), REGARDING HAVING LEAKS FROM THE HYDRO SYSTEM ON IMMAGE ANALYZER. NO ONE IN THE LABORATORY WAS AFFECTED OR EXPOSED BY THE LEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMMAGE® IMMUNOCHEMISTRY SYSTEM | NEPHELOMETER/IMMUNOCHEMISTRY SYSTEM | JQX | BECKMAN COULTER INC. | IMMAGE |