PROMUS EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2024168-2011-00702
- Event Type
- Malfunction
- Date Received
- February 4, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 11, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NIQ
- PMA / PMN Number
- P070015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. A FOLLOW-UP WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. YOU ARE RECEIVING THIS MDR REPORT FROM ABBOTT VASCULAR BECAUSE BOSTON SCIENTIFIC CORPORATION DISTRIBUTES PROMUS AS ITS OWN BRAND LABELING OF ABBOTT VASCULARS DRUG ELUTING STENT IN THE US. THE PROMUS 2.5 X 12 MM (PART#1009539-12B/LOT#0042941) IS BEING FILED UNDER A SEPARATE MANUFACTURER REPORT NUMBER.
(B)(4). THE INABILITY TO CROSS A LESION CAN BE AFFECTED BY NUMEROUS FACTORS INCLUDING, BUT NOT LIMITED TO, PATIENT ANATOMICAL MORPHOLOGY, PATIENT DISEASE STATE, PRE-DILATATION STRATEGY, DEVICE PLACEMENT TECHNIQUE, AND ACCESSORY DEVICE SUPPORT. IN THIS CASE, IT IS LIKELY THAT INTERACTION WITH THE HEAVILY CALCIFIED LESION CONTRIBUTED TO THE REPORTED FAILURE TO ADVANCE. ADDITIONALLY, AN INTERACTION WITH THE LESION/ANATOMY DURING THE ATTEMPT TO CROSS THE LESION MAY HAVE RESULTED IN AN INTERACTION WITH THE STENT IMPLANT THAT COULD HAVE CAUSED DAMAGE AND LOOSENED THE STENT ON THE BALLOON. A REVIEW OF THE PRODUCT MANUFACTURING RECORDS DID NOT REVEAL ANY NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT AND ALL LOT RELEASE TESTING MET MANUFACTURING CRITERIA. THE REPORTED FAILURE TO ADVANCE AND LOOSE STENT APPEAR TO BE RELATED TO THE OPERATIONAL CONTEXT OF THE PROCEDURE. ALL STENT DELIVERY SYSTEMS ARE INSPECTED FOR PROPER STENT PLACEMENT, STENT DAMAGE AND CRIMPED STENT OUTER DIAMETER. ADDITIONALLY, A SAMPLING OF UNITS IS DESTRUCTIVELY TESTED PRIOR TO RELEASE TO VERIFY STENT DISLODGEMENT FORCE.
IT WAS REPORTED THAT THE LESION WAS IN THE HEAVILY CALCIFIED RIGHT CORONARY ARTERY WITH MODERATE TORTOUSITY. BOTH RX PROMUS STENTS (3.0 X 18 MM AND 2.5 X 12 MM EXPERIENCED RESISTANCE WITH THE HEAVILY CALCIFIED LESION AND EACH DEVICE WAS REMOVED FROM THE ANATOMY. OUTSIDE THE PATIENT, THE STENTS WERE NOTED TO BE LOOSE. NO PATIENT EFFECTS WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROMUS EVEROLIMUS ELUTING CORONARY STENT SYSTEM | DRUG ELUTING CORONARY STENT SYSTEM | NIQ | AV-TEMECULA-CT | 0070142 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PROMUS 2.5 X 12 MM (PART#1009539-12B/LOT#0042941) |