LENSX LASER SYSTEM
Report
- Report Number
- 2028159-2024-01069
- Event Type
- Injury
- Date Received
- July 22, 2024
- Date of Event
- June 26, 2024
- Report Date
- October 18, 2024
- Manufacturer
- ALCON LENSX, INC.
- Product Code
- OOE
- UDI-DI
- 00380659981623
- PMA / PMN Number
- K163551
- Removal / Correction Number
- NA
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION PROVIDED IN D.9., H.3., H.6., AND H.11. A MANUFACTURING DEVICE HISTORY RECORD REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. THE COMPANY REPRESENTATIVE WAS UNABLE TO CONFIRM NOR REPLICATE THE REPORTED EVENT. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. THE CUSTOMER REPORTED EVENT CANNOT BE CONFIRMED. THUS, BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
H.3., H.6.: INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ALCON LENSX (SITE #3008772169) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER SITE #2028159). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A CUSTOMER REPORTED THAT THE MACHINE FAILED TO COMPLETELY CUT THE PATIENT'S MAIN AND SIDE PORTS IN AN UNSPECIFIED EYE, DURING A FLAP CREATION PROCEDURE. THE INCOMPLETE INCISIONS HAD TO BE MANUALLY COMPLETED WITH A KNIFE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1119418 | LENSX LASER SYSTEM | OPHTHALMIC FEMTOSECOND LASER | OOE | ALCON LENSX, INC. | 550 | NA | 00380659981623 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |