FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,10

MDR report key: 19790217 · Received July 22, 2024

Report

Report Number
0002023141-2024-02474
Event Type
Injury
Date Received
July 22, 2024
Date of Event
May 15, 2024
Report Date
December 6, 2024
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019225
PMA / PMN Number
K072589
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4) . A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. B5: DESCRIBE EVENT OR PROBLEM. G3: DATE RECEIVED BY MANUFACTURER. G6: TYPE OF REPORT. H1: TYPE OF REPORTABLE EVENT. H2: FOLLOW UP TYPE. H3: DEVICE EVALUATED BY MANUFACTURER. H6: ADVERSE EVENT PROBLEM. H10: ADDITIONAL NARRATIVE. ZIMVIE RECEIVED ONE (1) TSV4B10, (IMP,TSV,4.1MM,SBM,10) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE. IMPLANT FRACTURE IDENTIFIED AT THE COLLAR. ALSO, FRACTURED SCREW IDENTIFIED INSIDE THE IMPLANT. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1233732. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1233732 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: FRACTURE IMPLANT BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION ARE LONG-TERM PARAFUNCTIONAL HABITS (E.G., CLENCHING, BRUXISM, AND OVERLOADING) OF THE PATIENT OVER THE IMPLANTATION PERIOD PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

DOCTOR REPORTED CROWN MOBILITY DUE TO IMPLANT FRACTURE #27 AND WAS REMOVED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1120188 IMP,TSV,4.1MM,SBM,10 DENTAL IMPLANT DZE ZIMMER DENTAL 1233732 00889024019225

Patients

Seq Age Sex Outcome Treatment
1 70 YR Male