FDA Adverse Event Injury Summary report: N

STEM: AMISTEM P AMISTEM-P LAT STEM SIZE 3

MDR report key: 19790171 · Received July 22, 2024

Report

Report Number
3005180920-2024-00522
Event Type
Injury
Date Received
July 22, 2024
Date of Event
June 20, 2024
Report Date
July 19, 2024
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630040720120
PMA / PMN Number
K173794
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 27-JUN-2024. LOT 2214524: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 04-NOV-2022. EXPIRATION DATE: 2027-10-20. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DEPARTMENT: LATE INFECTION IN CEMENTLESS THA, 1.3 YEARS AFTER PRIMARY OPERATION. INFECTION IS A KNOWN POSSIBLE ADVERSE EVENT FOLLOWING EVERY SURGERY, INCLUDING THA'S. TO DATE, THERE IS NO REASON TO SUSPECT THAT THE CAUSE MAY BE LINKED TO THE IMPLANTED DEVICES. OTHER DEVICES INVOLVED: BATCH REVIEW PERFORMED ON 27-JUN-2024. LINER: MECTACER 01.29.414 CERAMIC LINER Ø 36 / F (NOT DISTRIBUTED IN US) LOT. 2217263 : (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25-OCT-2022. EXPIRATION DATE: 2027-10-04. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BALL HEADS: MECTACER 01.29.209 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE M 0 (K112115) LOT 2247815: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05-JAN-2023. EXPIRATION DATE: 2027-12-06. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL THE ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CUP: VERSAFITCUP CC TRIO 01.26.45.0056 ACETABULAR SHELL CC TRIO Ø 56 (K103352) LOT 2240318: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05-JAN-2023. EXPIRATION DATE: 2027-12-14. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

AT ABOUT 1 YEAR AND 3 MONTHS AFTER PRIMARY, THE PATIENT HAS BEEN REVISED BECAUSE OF INFECTION WHICH CAUSED STEM LOOSENING. NO MOBILIZATION OF THE ACETABULAR SHELL HAS BEEN DETECTED. THE SURGEON REVISED SUCCESSFULLY ALL COMPONENTS AND IMPLANTED A M-VIZION STEM WITH VERSAFITCUP DM COMPONENTS AND CERAMIC BALL HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1979708 STEM: AMISTEM P AMISTEM-P LAT STEM SIZE 3 HIP FEMORAL STEM LZO MEDACTA INTERNATIONAL SA 2214524 07630040720120

Patients

Seq Age Sex Outcome Treatment
1 53 YR Male Required Intervention