FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19789944 · Received July 22, 2024

Report

Report Number
9610877-2024-55589
Event Type
Malfunction
Date Received
July 22, 2024
Date of Event
July 10, 2024
Report Date
July 22, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
CAL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CFB (COHERENT FIBER BUNDLE). IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE CONTROL BODY FLUID DAMAGE, THE FORWARD BODY FRAME FLUID DAMAGE, THE INSERTION FLEXIBLE TUBE COATING DAMAGE, THE FORWARD BODY COVER BROKEN, AND THE SIDE BODY COVER BROKEN; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. THIS DEFECT OCCURRED NOT DURING PROCEDURE. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED NOT TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2249455 PENTAX FIBEROPTICINTUBATIONSCOPE 1.4C 3.4TP ROP CAL HOYA CORPORATION PENTAX TOKYO OFFICE FHY-10RBS

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown