Report Number
8021545-2024-02395
Event Type
Malfunction
Date Received
July 21, 2024
Date of Event
June 20, 2024
Report Date
December 16, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244023161
PMA / PMN Number
K210544
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 02 - MDR (B)(4) - MDR 8021545-2024-02395. CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT SUBMITTED DEVICE MANUFACTURER DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. CAPA-2010953 "TREND OBSERVED: INCREASE IN COMPLAINTS RELATED TO ADHESIVE ON EWIS" HAS BEEN OPENED ON 18-SEP-2024 TO ADDRESS ALL ADHESIVE ISSUES RELATED TO EWIS PRODUCT FAMILY AS NO SPECIFICATIONS EXIST FOR THE ADHESIVE USED ON THIS PRODUCT (DESIGN RELATED CAPA).

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4) - DEVICE 1 OF 6.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - MDR 1863725 - MDR 3003442380-2024-02395. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: D9: DEVICE AVAILABLE FOR EVALUATION HAS BEEN SELECTED AS YES & DATE RETURNED TO MFG WAS ADDED AS WELL. H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. ON (B)(6) 2024, PATIENT COMPLAINED ABOUT 6 INFUSION SETS FELL OFF. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1004561 MEDTRONIC EXTENDED INFUSION SET UNO EWIS BLUE 60/6 HCAP 3-PK INT FPA UNOMEDICAL A/S MMT-431AH 6005364 05705244023161

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown