INMODE RF
Report
- Report Number
- 3010511300-2024-00575
- Event Type
- Injury
- Date Received
- July 21, 2024
- Date of Event
- June 10, 2024
- Report Date
- August 19, 2024
- Manufacturer
- INMODE LTD.
- Product Code
- GEI
- UDI-DI
- 07290016633016
- PMA / PMN Number
- K151793
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
NO TECHNICAL ISSUES WERE FOUND IN THE SYSTEM DURING SERVICE INSPECTION. HP WAS NOT RETURNED FOR INSPECTION. THE INVESTIGATION ATTRIBUTED THE ROOT CAUSE TO USER ERRORS: CUT OFF TEMPERATURES USED WERE ABOVE THE RECOMMENDED PER IFU AND OBVIOUSLY TOO HIGH FOR THE GIVEN THIN FAT LAYER. THE TECHNIQUE ALSO APPEARS TO BE INCORRECT:POOR CONTROL OF TREATMENT DEPTH AND OF THE HP MOVEMENT. THE DOCTOR DECLINED A RETRAINING.
NO TECHNICAL ISSUES WERE FOUND IN THE SYSTEM DURING SERVICE INSPECTION. HP WAS NOT RETURNED FOR INSPECTION. THE INVESTIGATION ATTRIBUTED THE ROOT CAUSE TO USER ERRORS: CUT OFF TEMPERATURES USED WERE ABOVE THE RECOMMENDED PER IFU AND OBVIOUSLY TOO HIGH FOR THE GIVEN THIN FAT LAYER. THE TECHNIQUE ALSO APPEARS TO BE INCORRECT: POOR CONTROL OF TREATMENT DEPTH AND OF THE HP MOVEMENT. THE DOCTOR DECLINED A RETRAINING. (B)(6) 2024: CORRECTION IS SUBMITTED TO CORRECT H10 (MANUFACTURER NARRATIVE). HANDPIECE NAME WAS ADDED. THE HANDPIECE INVOLVED IN THE INCIDENT WAS FACETITE. NO TECHNICAL ISSUES WERE FOUND IN THE SYSTEM DURING SERVICE INSPECTION. HP WAS NOT RETURNED FOR INSPECTION. THE INVESTIGATION ATTRIBUTED THE ROOT CAUSE TO USER ERRORS: CUT OFF TEMPERATURES USED WERE ABOVE THE RECOMMENDED PER IFU AND OBVIOUSLY TOO HIGH FOR THE GIVEN THIN FAT LAYER. THE TECHNIQUE ALSO APPEARS TO BE INCORRECT: POOR CONTROL OF TREATMENT DEPTH AND OF THE HP MOVEMENT. THE DOCTOR DECLINED A RETRAINING.
FULL THICKNESS BURN ON JOWL.
FULL THICKNESS BURN ON JOWL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2483282 | INMODE RF | GEI | GEI | INMODE LTD. | AG604881A | 07290016633016 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Female | Other |