FDA Adverse Event Injury Summary report: N

INMODE RF

MDR report key: 19788435 · Received July 21, 2024

Report

Report Number
3010511300-2024-00575
Event Type
Injury
Date Received
July 21, 2024
Date of Event
June 10, 2024
Report Date
August 19, 2024
Manufacturer
INMODE LTD.
Product Code
GEI
UDI-DI
07290016633016
PMA / PMN Number
K151793
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

NO TECHNICAL ISSUES WERE FOUND IN THE SYSTEM DURING SERVICE INSPECTION. HP WAS NOT RETURNED FOR INSPECTION. THE INVESTIGATION ATTRIBUTED THE ROOT CAUSE TO USER ERRORS: CUT OFF TEMPERATURES USED WERE ABOVE THE RECOMMENDED PER IFU AND OBVIOUSLY TOO HIGH FOR THE GIVEN THIN FAT LAYER. THE TECHNIQUE ALSO APPEARS TO BE INCORRECT:POOR CONTROL OF TREATMENT DEPTH AND OF THE HP MOVEMENT. THE DOCTOR DECLINED A RETRAINING.

Additional Manufacturer Narrative · 0

NO TECHNICAL ISSUES WERE FOUND IN THE SYSTEM DURING SERVICE INSPECTION. HP WAS NOT RETURNED FOR INSPECTION. THE INVESTIGATION ATTRIBUTED THE ROOT CAUSE TO USER ERRORS: CUT OFF TEMPERATURES USED WERE ABOVE THE RECOMMENDED PER IFU AND OBVIOUSLY TOO HIGH FOR THE GIVEN THIN FAT LAYER. THE TECHNIQUE ALSO APPEARS TO BE INCORRECT: POOR CONTROL OF TREATMENT DEPTH AND OF THE HP MOVEMENT. THE DOCTOR DECLINED A RETRAINING. (B)(6) 2024: CORRECTION IS SUBMITTED TO CORRECT H10 (MANUFACTURER NARRATIVE). HANDPIECE NAME WAS ADDED. THE HANDPIECE INVOLVED IN THE INCIDENT WAS FACETITE. NO TECHNICAL ISSUES WERE FOUND IN THE SYSTEM DURING SERVICE INSPECTION. HP WAS NOT RETURNED FOR INSPECTION. THE INVESTIGATION ATTRIBUTED THE ROOT CAUSE TO USER ERRORS: CUT OFF TEMPERATURES USED WERE ABOVE THE RECOMMENDED PER IFU AND OBVIOUSLY TOO HIGH FOR THE GIVEN THIN FAT LAYER. THE TECHNIQUE ALSO APPEARS TO BE INCORRECT: POOR CONTROL OF TREATMENT DEPTH AND OF THE HP MOVEMENT. THE DOCTOR DECLINED A RETRAINING.

Description of Event or Problem · 0

FULL THICKNESS BURN ON JOWL.

Description of Event or Problem · 0

FULL THICKNESS BURN ON JOWL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2483282 INMODE RF GEI GEI INMODE LTD. AG604881A 07290016633016

Patients

Seq Age Sex Outcome Treatment
1 57 YR Female Other