FDA Adverse Event Injury Summary report: N

MINIMED QUICK-SET

MDR report key: 19786394 · Received July 20, 2024

Report

Report Number
3003442380-2024-15351
Event Type
Injury
Date Received
July 20, 2024
Date of Event
June 11, 2024
Report Date
July 31, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244017450
PMA / PMN Number
K160648
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PATIENT COUNTRY: (B)(6) PATIENT CITY: WARSZAWA.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE BATCH 6001733 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6001733 WAS MANUFACTURED ACCORDING TO THE WI VERSION 75 PACKAGING IN THE MULTIVAC 12, ON 07/JUN/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 30/JUL/2025 AGAINST MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION, HARM CODE UNTREATED DIABETIC KETOACIDOSIS WHICH THE PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY A HEALTH CARE PROVIDER (HCP) OR REQUIRES EMERGENCY ADVANCE, AND LOT 6001733 AND NO OTHER MORE COMPLAINT HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT NUMBER AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION AND HARM CODE, AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION OR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN POLAND. IT WAS REPORTED THAT PATIENT FACED AN INFUSION SET HAD AIR BUBBLE IN THE TUBING AND WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE, DIABETES KETOACIDOSIS AND COMA ON (B)(6) 2024. THE BLOOD GLUCOSE LEVEL WAS 400 MG/DL AT THE TIME OF EVENT AND WAS MANAGED BY INSULIN PEN. THE LENGTH OF HOSPITALIZATION WAS 5 DAYS. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1918878 MINIMED QUICK-SET UNO QUICK-SET 60/6 SC1 MECA FPA UNOMEDICAL A/S MMT-399A 6001733 05705244017450

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| H