BEAVER-VISITEC INTERNATIONAL INC
Report
- Report Number
- 1211998-2024-00014
- Event Type
- Malfunction
- Date Received
- July 19, 2024
- Date of Event
- July 9, 2024
- Report Date
- October 21, 2024
- Manufacturer
- BEAVER-VISITEC INTERNATIONAL INC
- Product Code
- HNN
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
BVI QUALITY ASSURANCE HAS FOLLOWED ITS INTERNAL PROCEDURE TO CONDUCT THE APPROPRIATE COMPLAINT INVESTIGATION. THE INVESTIGATION INCLUDES A REVIEW OF THE CURRENT PROCESS AND DEVICE HISTORY RECORD (DHR). ALL REQUIRED ELEMENTS OF THE DHR WERE PROPERLY FILLED. ALL MANUFACTURING OPERATIONS WERE RECORDED; ALL SIGNATURES AND DATES WERE PRESENT, AND THE LOT PASSED THE REQUIRED INSPECTIONS. THE CUSTOMER PROVIDED PICTURES OF THE DEFECTIVE PRODUCT FOR FURTHER INVESTIGATION. UPON REVIEWE, THE REPORTED DEFECT "SHIELD RETRACTED" ON THE KNIFE'S PACKAGING WAS CLEARLY VISIBLE AND CONFIRMED. ACCORDING TO THE INVESTIGATION, THE DEFECT MAY APPEAR DURING TRANSPORTATION CAUSING THE SHIELD TO RETRACT. WHICH IS NOT ACCEPTABLE ACCORDING TO PRODUCT SPECIFICATIONS. BVI WILL CONTINUE TO MONITOR ANY FEEDBACK FROM OUR CUSTOMERS AND TAKE APPROPRIATE ACTION IF AN UNFAVOURABLE TREND IS OBSERVED.
BVI HAS BEEN NOTIFIED ABOUT KNIFES THAT WERE PACKED WITHOUT THE PROTECTION ON THE BLADE AND THERE IS ONE INCIDENT OF A NURSE "STABBING" HERSELF WHILE HANDLING THE KNIFE STILL PACKED IN THE WHITE CONTAINER. THE INCIDENT OCCURRED IN NETHERLANDS.
THE HOSPITAL PROVIDED ADDITIONAL INFORMATION REGARDING THE NURSE OUTCOME. AS IT WAS STATED, THE NURSING STAFF DID NOT REPORT MISSED WORKING HOURS AND THERE WAS NO INTERNAL REPORT OF A NEEDLESTICK INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1552368 | BEAVER-VISITEC INTERNATIONAL INC | BEAVER® SAFETY SIDEPORT KNIFE 1.15MM 45° | HNN | BEAVER-VISITEC INTERNATIONAL INC | 378231 | 6066482 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |