FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE¿

MDR report key: 19784630 · Received July 19, 2024

Report

Report Number
2024168-2024-08599
Event Type
Injury
Date Received
July 19, 2024
Date of Event
June 21, 2024
Report Date
September 24, 2024
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648113154
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Additional Manufacturer Narrative · 0

AN ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE FAILURE TO CYCLE WAS CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED ISSUE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON OBSERVATIONS FROM THE RETURNED ANALYSIS A POSTERIOR NEEDLE TO CUFF MISS OCCURRED. THE POSTERIOR NEEDLE TIP APPEARS TO HAVE STRIPPED OF THE SUTURE DURING REMOVAL. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4 CORRECTION: LOT NUMBER UPDATED FROM 3110741 TO 3102641.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE LEFT COMMON FEMORAL ARTERY USING A PROGLIDE DEVICE AFTER AN ARTERIOVENOUS MALFORMATION (AVM) EMBOLIZATION WITH A 6F SHEATH. REPORTEDLY, THE SUTURE DID NOT COME OUT WHEN THE HANDLE [PLUNGER] WAS RETRACTED. A NEW PROGLIDE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1754762 PERCLOSE PROGLIDE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12673-05 3102641 08717648113154

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention