PERCLOSE PROGLIDE¿
Report
- Report Number
- 2024168-2024-08599
- Event Type
- Injury
- Date Received
- July 19, 2024
- Date of Event
- June 21, 2024
- Report Date
- September 24, 2024
- Manufacturer
- ABBOTT VASCULAR INC.
- Product Code
- MGB
- UDI-DI
- 08717648113154
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.
AN ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE FAILURE TO CYCLE WAS CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED ISSUE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON OBSERVATIONS FROM THE RETURNED ANALYSIS A POSTERIOR NEEDLE TO CUFF MISS OCCURRED. THE POSTERIOR NEEDLE TIP APPEARS TO HAVE STRIPPED OF THE SUTURE DURING REMOVAL. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4 CORRECTION: LOT NUMBER UPDATED FROM 3110741 TO 3102641.
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE LEFT COMMON FEMORAL ARTERY USING A PROGLIDE DEVICE AFTER AN ARTERIOVENOUS MALFORMATION (AVM) EMBOLIZATION WITH A 6F SHEATH. REPORTEDLY, THE SUTURE DID NOT COME OUT WHEN THE HANDLE [PLUNGER] WAS RETRACTED. A NEW PROGLIDE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1754762 | PERCLOSE PROGLIDE¿ | DEVICE, HEMOSTASIS, VASCULAR | MGB | ABBOTT VASCULAR INC. | 12673-05 | 3102641 | 08717648113154 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | Required Intervention |