FDA Adverse Event Death Summary report: N

BIPAP AUTOSV ADV

MDR report key: 19780431 · Received July 19, 2024

Report

Report Number
2518422-2024-45850
Event Type
Death
Date Received
July 19, 2024
Date of Event
March 6, 2023
Report Date
July 11, 2025
Manufacturer
RESPIRONICS, INC.
Product Code
MNS
UDI-DI
00606959012530
PMA / PMN Number
K090248
Removal / Correction Number
Z-1973-2021
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

IN PREVIOUS REPORT DESCRIBE EVENT OR PROBLEM WAS CAPTURED INCORRECTLY AND IT HAS BEEN CORRECTED IN THIS REPORT. THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A BIPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED DEATH. THE DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER FOR FURTHER EVALUATION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND NO EVIDENCE OF FOAM DEGRADATION. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WAS NO ERROR CODE FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY COULDN'T CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS NO VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED. IN BOX B: ADVERSE EVENT/PRODUCT PROBLEM UPDATED. IN BOX D: MODEL NUMBER, LOT # (RFB), CATALOG ITEM IDENTIFIER, EXPIRATION DATE (RFB) AND PRODUCT UDI CORRECTED. DEVICE SERVICED BY 3RDP?, DEVICE AVAILABLE FOR EVAL? HAS BEEN UPDATED. DATE RECEIVED BY MFG, REPORT TYPE, 510K AND IN BOX H, IF FOLLOW-UP, WHAT TYPE?, DEVICE PROBLEM CODE GRID, EVALUATION METHOD CODE GRID, EVALUATION RESULTS CODE GRID CONCLUSION CODE GRID, REMEDIAL ACTION INITIATED AND RECALL (Z) NUMBER HAS BEEN UPDATED IN THIS REPORT.

Description of Event or Problem · 0

THE MANUFACTURER WAS CONTACTED IN REFERENCE TO A BIPAP AUTOSV ADV DEVICE. THE PATIENT PASSED AWAY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1555982 BIPAP AUTOSV ADV VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING MNS RESPIRONICS, INC. DS960TS 00606959012530

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Death