FDA Adverse Event Injury Summary report: N

NOVA MAX

MDR report key: 1977753 · Received January 28, 2011

Report

Report Number
MW5019213
Event Type
Injury
Date Received
January 28, 2011
Date of Event
November 20, 2010
Report Date
January 28, 2011
Manufacturer
NOVA BIOMEDICAL
Product Code
NBW
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IL, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

I WOULD LIKE TO BRING TO YOUR ATTENTION FALSE READINGS BY YOUR NOVA MAX GLUCOSE TEST STRIPS. LOT NUMBER 1020210181 EXP DATE 2012-06, PART NUMBER: 43437. SAT (B)(6) 2010: I SPENT 6 HOURS GIVING MYSELF 110 UNITS OF INSULIN AS A RESULT OF INACCURATE TEST STRIP READINGS. I ENDED UP IN THE EMERGENCY ROOM WITH A SUGAR LEVEL OF 24, COMPLETELY INCOHERENT AND ON THE VERGE OF A DIABETIC COMA. ONCE THE DOCTORS REVIVED AND STABILIZED ME, WE COMPARED SUGAR LEVELS BETWEEN BY MACHINE WITH YOUR TEST STRIPS AND THE HOSPITAL'S MACHINES. THE TEST STRIPS SHOWED MY SUGAR LEVELS RANGING FROM 555-HIGH WHEN COMPARED TO THE EMERGENCY ROOM MACHINES READINGS OF 140. THIS SITUATION NEARLY COST ME MY LIFE AND ONCE I CHANGED MY TEST STRIPS, I FINALLY OBTAINED ACCURATE READINGS AND WAS DISCHARGED FROM THE HOSPITAL. DATES OF USE: (B)(6) 2010.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NOVA MAX GLUCOSE TEST STRIPS, 50 NBW NOVA BIOMEDICAL 1020210181

Patients

Seq Age Sex Outcome Treatment
1 38 YR Hospitalization| L