SURESCAN
Report
- Report Number
- 3004209178-2024-15038
- Event Type
- Malfunction
- Date Received
- July 18, 2024
- Date of Event
- July 1, 2024
- Report Date
- July 18, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LGW
- UDI-DI
- 00643169109483
- PMA / PMN Number
- P840001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR SPINAL PAIN. IT WAS REPORTED THAT THIS MORNING THEY STARTED SEEING A POR (POWER ON RESET) MESSAGE ON THEIR RECHARGER WITH A CAUTION SIGN. CALLER STATED THEY COULDN'T SEEM TO CLEAR THE MESSAGE. AGENT HAD CALLER CONNECT THE PROGRAMMER TO THE IMPLANT. CALLER CONFIRMED SEEING THE INFORMATIONAL POR MESSAGE ON THE PROGRAMMER. AGENT WALKED CALLER THROUGH CLEARING THE POR MESSAGE. CALLER CONFIRMED THE POR POWER ON RESET MESSAGE CLEARED, AND THEY WERE ABLE TO ACCESS THE THERAPY SCREEN. AGENT REVIEWED WITH CALLER EVERYTHING APPEARS TO BE WORKING AS INTENDED AND INSTRUCTED TO MONITOR THE ISSUE AND CALL BACK IF IT PERSISTS. THE TROUBLESHOOTING STEPS THAT WERE TAKEN ON THE CALL RESOLVED THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1918298 | SURESCAN | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR | LGW | MEDTRONIC PUERTO RICO OPERATIONS CO. | 97714 | 00643169109483 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Female |