FDA Adverse Event Other Summary report: N

MULTISCAN TABLE MODEL 8083

MDR report key: 1977266 · Received January 28, 2011

Report

Report Number
1932056-2011-00001
Event Type
Other
Date Received
January 28, 2011
Date of Event
November 9, 2009
Report Date
January 25, 2011
Manufacturer
MEDICAL POSITIONING, INC.
Product Code
LGX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INCIDENT RESULTED FROM WELD FAILURE. WHILE LESS THAN 0.02% WELDS HAVE FAILED, THE INCIDENCE OF FAILURE IN 2004 IS HIGHER THAN IN OTHER YEARS. FOR THAT REASON, ALL CUSTOMERS WITH MODELS WHO HAVE REPORTED WELD FAILURE, AND ARE FROM BATCHES OF STEEL USED IN 2004, HAVE BEEN CONTACTED AND AN UPGRADE WAS INSTALLED TO ENSURE NO FAILURES IN THE FUTURE.

Description of Event or Problem · 1

WITH THE PT ON THE TABLE IN THE FOWLER POSITION (SITTING UP WITH BACK BEING SUPPORTED BY THE UPPER PORTION OF THE TABLE), THE LOWER FOWLER ACTUATOR MOUNT TAB BROKE AWAY FROM THE TABLE CAUSING THE HEAD OF THE TABLE TO FALL TO THE FLOOR. THE PT WAS ABLE TO AVOID INJURY BY UTILIZING THE SAFETY MECHANISMS BUILT INTO THE TABLE. SPECIFICALLY, THE PT GRABBED ONTO THE SAFETY RAIL BEFORE HE FELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MULTISCAN TABLE MODEL 8083 GENERAL ULTRASOUND TABLE LGX MEDICAL POSITIONING, INC. 8083

Patients

Seq Age Sex Outcome Treatment
1 Other