BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE
Report
- Report Number
- 1917413-2024-00653
- Event Type
- Malfunction
- Date Received
- July 18, 2024
- Date of Event
- June 4, 2024
- Report Date
- August 14, 2024
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JCF
- UDI-DI
- 30382903627616
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D4. MEDICAL DEVICE LOT#: UNKNOWN. D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. ADDITIONALLY, BD WAS UNABLE TO DETERMINE THE SPECIFIC LOT NUMBER ASSOCIATED WITH THIS COMPLAINT; THEREFORE, A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED. THIS COMPLAINT IS UNABLE TO BE CONFIRMED. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THIS COMPLAINT WILL BE REOPENED TO ASSESS THE LEVEL OF INVESTIGATION NEEDED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE AN UNSPECIFIED NUMBER OF TUBES BROKE IN THE CENTRIFUGE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE AN UNSPECIFIED NUMBER OF TUBES BROKE IN THE CENTRIFUGE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1467664 | BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE | LYMPHOCYTE SEPARATION MEDIUM | JCF | BECTON, DICKINSON & CO. (BROKEN BOW) | UNKNOWN | 30382903627616 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |