FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 19771684 · Received July 18, 2024

Report

Report Number
1000113657-2024-00298
Event Type
Malfunction
Date Received
July 18, 2024
Date of Event
June 26, 2024
Report Date
July 18, 2024
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER WAS RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND NO DEFECT FOUND ON RETURNED METER. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-062: USER HAD POOR TECHNIQUE. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 03-JUL-2024 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 116, 119, 142, 114 AND 132 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT IS BELOW 80 MG/DL AND EXPECTED PM FASTING BLOOD GLUCOSE TEST RESULT IS BELOW 120 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK-TO-BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 100 MG/DL AND 96 MG/DL USING TRUE METRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE OFFICE. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 09/30/2025 AND TEST STRIPS WERE OPENED 3 WEEKS PRIOR TO THE CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 114 MG/DL DATE: 6/26/2024 TIME: 9:29 AM FASTING RESULT 2: 132 MG/DL DATE: 6/26/2024 TIME: 9:26 AM FASTING RESULT 3: 119 MG/DL DATE: 6/26/2024 TIME: 9:26 AM FASTING RESULT 4: 137 MG/DL DATE: 6/26/2024 TIME: 9:16 AM FASTING RESULT 5: 125 MG/DL DATE: 6/26/2024 TIME: 9:15 AM FASTING RESULT 6: 116 MG/DL DATE: 6/25/2024 TIME: 8:36 PM FASTING RESULT 7: 119 MG/DL DATE: 6/25/2024 TIME: 8:36 PM FASTING RESULT 8: 142 MG/DL DATE: 6/25/2024 TIME: 8:34 PM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1551384 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX SGC 100CT12/CS MG/DL ZC5615S

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown