FDA Adverse Event Other Summary report: N

DA VINCI SURGCAL SYSTEM

MDR report key: 1977151 · Received February 2, 2011

Report

Report Number
2955842-2011-00033
Event Type
Other
Date Received
February 2, 2011
Date of Event
January 3, 2011
Report Date
January 3, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K011002
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING CONCLUDED THAT THE SYSTEM ISSUE EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH THE LEFT MASTER TOOL MANIPULATOR (MTML) GIMBAL. THE MTM GIMBAL IS A SUBSECTION OF THE COMPLETE MTM ARM, CONSISTING OF THE LINKS ASSOCIATED WITH THE FIVE AXES OF MOTION CLOSEST TO THE SURGEON'S HAND. THESE AXES MEASURE THE ORIENTATION OF THE MTM HANDLE AND THE MTM GRIP ANGLE. THE GIMBAL IS SEPARATED FROM THE REST OF THE MTM BY REMOVING THE PLATFORM LINK FROM THE FOREARM LINK. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED MTML GIMBAL. THE MTML GIMBAL WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING EVALUATION OF THE MTML GIMBAL FOUND THE GIMBAL GRIP LEVERS HAD EXCESSIVE PLAY, THUS CONFIRMING THE REPORTED COMPLAINT EXPERIENCED BY THE CUSTOMER. THE GIMBAL GRIP LEVERS WERE REPLACED. AS OF (B)(4), 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SACROCOLPOPEXY PROCEDURE, THE SURGEON EXPERIENCED DIFFICULTY WITH THE GRIPS OPENING AND CLOSING FOR THE INSTRUMENTS CONTROLLED BY THE LEFT MASTER TOOL MANIPULATOR (MTM). THE PATIENT HAD BEEN UNDER ANESTHESIA FOR APPROXIMATELY ONE HOUR AND 45 MINUTES WHEN THE SURGEON DECIDED TO CONVERT TO OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED PROCEDURE. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI SURGCAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS1200 A4.3P9

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI SYSTEM INSTRUMENTS AND ACCESSORIES