FDA Adverse Event Malfunction Summary report: N

PEN NDL 32G 4MM PRO 100 BOX 1200 US

MDR report key: 19754658 · Received July 16, 2024

Report

Report Number
3023359743-2024-00296
Event Type
Malfunction
Date Received
July 16, 2024
Date of Event
June 21, 2024
Report Date
October 8, 2024
Manufacturer
EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
Product Code
FMI
UDI-DI
00382903205509
PMA / PMN Number
K212015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WAS ADDED TO: B4, G6, H2, H3, H11. CORRECTION TO: H6 (COMPONENT CODE, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS) INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AS BENT CANNULA NPE. BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.

Description of Event or Problem · 0

CONSUMER REPORTED, SHE WAS ATTEMPTING TO INJECT 14 UNITS OF INSULIN BUT THE FLOW STOPPED BEFORE SHE WAS COMPLETE. STATED, SHE HAD TO USE A SECOND PEN NEEDLE TO COMPLETE HER INJECTIONS. 1 PEN NEEDLE AFFECTED. LOT: 3241355. CATALOG: 320550. DATE OF EVENT: 2024-06-21. SAMPLES: YES CL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2524131 PEN NDL 32G 4MM PRO 100 BOX 1200 US NEEDLE, HYPODERMIC, SINGLE LUMEN FMI EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND 320550 3241355 00382903205509

Patients

Seq Age Sex Outcome Treatment
1 NA Female