FDA Adverse Event Malfunction Summary report: N

4.5 HEALIX ADVANCE BR W/OCORD

MDR report key: 19752783 · Received July 16, 2024

Report

Report Number
1221934-2024-02482
Event Type
Malfunction
Date Received
July 16, 2024
Date of Event
July 2, 2024
Manufacturer
DEPUY MITEK LLC US
Product Code
MAI
UDI-DI
10886705021314
PMA / PMN Number
K120078
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. H10 ADDITIONAL NARRATIVE: H4: THE DEVICE MANUFACTURE DATE IS CURRENTLY UNAVAILABLE. INVESTIGATION SUMMARY: THE PRODUCT HAS NOT RETURNED TO DEPUY SYNTHES MITEK, HOWEVER A PHOTO WAS PROVIDED FOR REVIEW. THE PHOTO INVESTIGATION REVEALED THAT 4.5 HEALIX ADVANCE BR W/OCORD ANCHOR WAS BROKEN AT THE DISTAL END. NO OTHER ANOMALIES COULD BE OBSERVED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF 4.5 HEALIX ADVANCE BR W/OCORD WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL CAUSE FOR THE BROKEN ANCHOR CAN BE TRACED TO OFF AXIS INSERTION AND LEVERING DURING INSERTION. AS PER IFU: INSERTING THE AWL OR DRILL TO LESS THAN THE SPECIFIED DEPTH, AXIAL MISALIGNMENT OR LEVERING WITH THE ANCHOR UPON INSERTION MAY RESULT IN ANCHOR FRACTURE, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES MITEK QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. H10 ADDITIONAL NARRATIVE: D9, H3, H6: THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. H10 ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY THE PRODUCT WAS RETURNED TO DEPUY SYNTHES MITEK FOR EVALUATION. THE DEPUY SYNTHES MITEK TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL INSPECTION FOUND THAT 4.5 HEALIX ADVANCE BR W/OCORD HAD THE ANCHOR STILL ATTACHED TO THE INSERTER AND BROKEN AT THE MIDDLE END. THE ANCHOR HAD FOREIGN MATTER, PRESUMABLY BIOLOGICAL MATTER AT THE BASE OF THE ANCHOR. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF 4.5 HEALIX ADVANCE BR W/OCORD WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE.¿ BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL CAUSE IS TRACED TO THE PROCEDURAL VARIABLES, SUCH HANDLING OF THE DEVICE OR PRODUCT INTERACTION DURING PROCEDURE. THE POSSIBLE ROOT CAUSE CAN BE ASSOCIATED WITH OFF AXIS INSERTION AND LEVERING DURING INSERTION. AS PER IFU-100191 AXIAL MISALIGNMENT OR LEVERING WITH THE ANCHOR UPON INSERTION MAY RESULT IN ANCHOR FRACTURE. AS THE DEVICE SHOW SIGNS OF USE, THE REPORTED COMPLAINT OF DEVICE DEFORMED UPON OPENING OF THE PACKAGE CANNOT BE CONFIRMED, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES MITEK QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

IT WAS REPORTED FROM CHINA THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, IT WAS OBSERVED THAT THE ANCHOR ON THE 4.5 HEALIX ADVANCE BR W/OCORD DEVICE BROKE OFF, THEN ALL THE BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2505146 4.5 HEALIX ADVANCE BR W/OCORD FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE MAI DEPUY MITEK LLC US 261L018 10886705021314

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown