Report
- Report Number
- 1527736-2011-00029
- Event Type
- Malfunction
- Date Received
- January 31, 2011
- Date of Event
- November 24, 2010
- Report Date
- December 22, 2010
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES NOTED. EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE WAS RECEIVED FOR ANALYSIS. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A PLEURODESE OR BIOPSY PROCEDURE, THE SURGEON FIRED, BUT THE STAPLER LOCKED. IT IS UNKNOWN HOW THE CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON*FLEX60 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO SURGERY, INC (CINCINNATI) | UNK | G4TNV3 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |