FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60

MDR report key: 1974452 · Received January 31, 2011

Report

Report Number
1527736-2011-00029
Event Type
Malfunction
Date Received
January 31, 2011
Date of Event
November 24, 2010
Report Date
December 22, 2010
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES NOTED. EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE WAS RECEIVED FOR ANALYSIS. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PLEURODESE OR BIOPSY PROCEDURE, THE SURGEON FIRED, BUT THE STAPLER LOCKED. IT IS UNKNOWN HOW THE CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 STAPLE, IMPLANTABLE GDW ETHICON ENDO SURGERY, INC (CINCINNATI) UNK G4TNV3

Patients

Seq Age Sex Outcome Treatment
1