FDA Adverse Event Malfunction Summary report: N

POSEY SITTER II ALARM

MDR report key: 1974098 · Received December 30, 2010

Report

Report Number
2020362-2010-00414
Event Type
Malfunction
Date Received
December 30, 2010
Report Date
October 19, 2010
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL: EVAL SHOWS THE RETURNED PRODUCT FAIL SAFE SOUNDS INTERMITTENTLY WHEN THE SENSOR IS UNPLUGGED. PIN # 4 OF SENSOR # 2 RECEPTACLE IS BENT. THERE'S DAMAGE TO THE PLASTIC MOLD OF THE SENSOR # 1 RECEPTACLE. THE ALARM CASE HAS DAMAGE NEAR THE BATTERY COMPARTMENT. NOTE: THE POSEY ALARM IS AN ELECTRONIC DEVICE THAT MAY FAIL TO WORK IF SUBJECTED TO SEVERE SHOCK, SUCH AS BEING DROPPED (BENT) OR IMMERSED IN LIQUID. THE UNIT MAY STOP FUNCTIONING AS DESIGNED. (B)(4).

Description of Event or Problem · 1

CUSTOMER CLAIMS THAT WHEN A SENSOR 8308 IS INSERTED INTO THE FIRST SENSOR RECEPTACLE THE ALARM SOUNDS CONTINUOUSLY WHEN WEIGHT IS APPLIED TO THE SENSOR. WHEN THE RJ-11 CLIP IS INSERTED, IT DOES FIT PROPERLY. THERE IS NO VISIBLE DAMAGE TO THE ALARM. THERE WAS NO PT INCIDENT OR INJURY REPORTED. EVAL SHOWS THE ALARM SOUNDS INTERMITTENTLY WHEN THE SENSOR IS UNPLUGGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POSEY SITTER II ALARM KMI J. T. POSEY CO. 8281 NA

Patients

Seq Age Sex Outcome Treatment
1 NI SQUARE CHAIR SENSOR PAD, MODEL # 8308, LOT # UNK