MEDTRONIC EXTENDED
Report
- Report Number
- 8021545-2024-02142
- Event Type
- Malfunction
- Date Received
- July 15, 2024
- Date of Event
- June 14, 2024
- Report Date
- December 17, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244022836
- PMA / PMN Number
- K210544
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
E1: PATIENT CITY: (B)(6), PATIENT COUNTRY: UNITED STATES.
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED DEVICE MANUFACTURER DATE UNDER H4. CAPA-2010953 "TREND OBSERVED: INCREASE IN COMPLAINTS RELATED TO ADHESIVE ON EWIS" HAS BEEN OPENED ON 18-SEP-2024 TO ADDRESS ALL ADHESIVE ISSUES RELATED TO EWIS PRODUCT FAMILY AS NO SPECIFICATIONS EXIST FOR THE ADHESIVE USED ON THIS PRODUCT (DESIGN RELATED CAPA).
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: CAPA-2010953 "TREND OBSERVED: INCREASE IN COMPLAINTS RELATED TO ADHESIVE ON EWIS" HAS BEEN OPENED ON 18-SEP-2024 TO ADDRESS ALL ADHESIVE ISSUES RELATED TO EWIS PRODUCT FAMILY AS NO SPECIFICATIONS EXIST FOR THE ADHESIVE USED ON THIS PRODUCT (DESIGN RELATED CAPA).
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES, IT WAS REPORTED THAT ON 14-JUN-2024 THE ONE INFUSION SET WAS FELL OFF DURING USE THE BLOOD GLUCOSE LEVEL WAS HIGH AND THE ISSUE IS DUE TO PATIENT LEADING TO HIGH BLOOD GLUCOSE RUNNING HIGH AND PATIENT TAKE A MANUAL INJECTION. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256688 | MEDTRONIC EXTENDED | UNO EWIS BLUE 80/9 HCAP 10PK INT | FPA | UNOMEDICAL A/S | MMT-442A | 6004032 | 05705244022836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |