FDA Adverse Event Malfunction Summary report: N

INTEGRATED APD SET W/CASSETTE3-PRONG

MDR report key: 1973940 · Received January 31, 2011

Report

Report Number
1423500-2011-01308
Event Type
Malfunction
Date Received
January 31, 2011
Date of Event
January 7, 2011
Report Date
January 7, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS USE ERROR. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER THE SAMPLE WAS DISCARDED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED GLOBAL TECHNICAL SERVICE (GTS) REGARDING A SYSTEM ERROR 2240 THAT APPEARED ON THE HOME CHOICE (HC) DURING INITIAL DRAIN. THE HOME PATIENT (HP) STATED THAT THE CLAMP ON THE PATIENT LINE WAS CLOSED AND WHEN HE CONNECTED, HE OPENED IT AND PRESSED GO TO START INITIAL DRAIN. THE HP THEN GOT THE SYSTEM ERROR 2240. GTS EXPLAINED THE ALARM TO THE HP AND THAT HE WOULD NEED TO SETUP WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. DURING A FOLLOW UP, WITH THE HOME PATIENT'S NURSE, SHE STATED THAT THE HP NOTIFIED HER AND THAT SHE SAW HIM TODAY. THE NURSE STATED THAT HE TOLD HER ALL ABOUT IT AND SAID IT WAS HIS FAULT. THE NURSE SAID EVERYTHING WAS FINE, AND THAT HE KNEW EXACTLY WHAT HE DID WRONG. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTEGRATED APD SET W/CASSETTE3-PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 77 YR HOMECHOICE CYCLER