FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 1973730 · Received January 17, 2011

Report

Report Number
1720753-2011-00325
Event Type
Malfunction
Date Received
January 17, 2011
Date of Event
December 21, 2010
Report Date
January 17, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE HIGH VOLTAGE CABLE WAS CLEANED AND REGREASED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9600 SYSTEM INTERMITTENTLY LOCKED UP AND HAS TO BE REBOOTED AND GOES BLACK DURING A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1