FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1973673 · Received January 18, 2011

Report

Report Number
1720753-2011-00349
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
December 21, 2010
Report Date
January 18, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CABLE WAS FOUND PULLED OUT OF THE PRINTER. THE PRINTER CABLE CONNECTION WAS REPAIRED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM LOCKED UP WHEN TRYING TO PRINT IMAGES OUTSIDE A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1