FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1973671 · Received January 18, 2011

Report

Report Number
1720753-2011-00357
Event Type
Malfunction
Date Received
January 18, 2011
Date of Event
December 21, 2010
Report Date
January 18, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE TRANSFORMER WAS ADJUSTED. THE PS1 POWER SUPPLY VOLTAGE WAS ADJUSTED AND THE INTERCONNECT CABLE AND LEMO CONNECTOR WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM HAD AN INTERMITTENT COMMUNICATION ERROR DURING A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1