FDA Adverse Event Malfunction Summary report: N

MEDTRONIC EXTENDED

MDR report key: 19736509 · Received July 13, 2024

Report

Report Number
8021545-2024-01988
Event Type
Malfunction
Date Received
July 13, 2024
Date of Event
June 12, 2024
Report Date
December 17, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244022836
PMA / PMN Number
K210544
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR 1911190 - MDR 8021545-2024-01988 - DEVICE 1 OF 2. E1: PATIENT CITY: (B)(4). PATIENT COUNTRY: THE UNITED STATES.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 8021545-2024-01988. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: "TREND OBSERVED: INCREASE IN COMPLAINTS RELATED TO ADHESIVE ON EWIS" HAS BEEN OPENED ON 18-SEP-2024 TO ADDRESS ALL ADHESIVE ISSUES RELATED TO EWIS PRODUCT FAMILY AS NO SPECIFICATIONS EXIST FOR THE ADHESIVE USED ON THIS PRODUCT (DESIGN RELATED CAPA).

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 02 - MDR (B)(4). ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. CAPA-2010953 "TREND OBSERVED: INCREASE IN COMPLAINTS RELATED TO ADHESIVE ON EWIS" HAS BEEN OPENED ON 18-SEP-2024 TO ADDRESS ALL ADHESIVE ISSUES RELATED TO EWIS PRODUCT FAMILY AS NO SPECIFICATIONS EXIST FOR THE ADHESIVE USED ON THIS PRODUCT (DESIGN RELATED CAPA).

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED PROBLEM WITH TWO INFUSION SETS FELL OFF DURING USE EVENTS ON 12-JUNE-2024. PATIENT REPORTED AN ISSUE WITH INFUSION SET TAPE STICKING. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
833650 MEDTRONIC EXTENDED UNO EWIS BLUE 80/9 HCAP 10PK INT FPA UNOMEDICAL A/S MMT-442A 6003819 05705244022836

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown