FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1973595 · Received January 19, 2011

Report

Report Number
1720753-2011-00395
Event Type
Malfunction
Date Received
January 19, 2011
Date of Event
December 23, 2010
Report Date
January 19, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE HIGH VOLTAGE CABLE WAS CLEANED AND REGREASED AND THE FILAMENT WAS CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM LOCKED UP WITH AN OVERLOAD FAULT ERROR OUTSIDE A CASE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1