PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED
Report
- Report Number
- 1423500-2011-01286
- Event Type
- Malfunction
- Date Received
- January 31, 2011
- Date of Event
- January 6, 2011
- Report Date
- January 6, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE IS NOT AVAILABLE FOR EVALUATION. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THEN A FOLLOW UP MDR WILL BE SUBMITTED.
(B)(4). THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE CAUSE FOR THE THERAPY SHORTENED FROM 5 TO 4 CYCLES WAS UNDETERMINED. THE DEVICE'S SERVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE DEVICE CHANGING THE CYCLES. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
A CUSTOMER CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING THE HOMECHOICE (HC) UNIT SHORTENING THE NUMBER OF CYCLES DURING USE. THE HOME PATIENT (HP) REPORTED THAT THE CYCLES IN HIS THERAPY HAD SHORTENED FROM 5 TO 4 CYCLES. HP WANTED TO KNOW IF IT IS POSSIBLE THAT THE MACHINE BYPASSES ON IT'S OWN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE POSSIBILITIES OF WHY CYCLES WOULD BE SHORTENED AND INSTRUCTED HP TO CALL BAXTER AT TIME OF THE EVENT TO TROUBLESHOOT. THERE WAS PATIENT INVOLVEMENT. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. PRODUCT SURVEILLANCE FOLLOWED UP WITH THE HP'S NURSE ON (B)(4) 2011 REGARDING THE REPORTED PROBLEM. THE NURSE STATED, HP WAS IN THE CLINIC YESTERDAY AND HE DID INFORM HER ABOUT THE ISSUE WITH THE HC SHORTENING THE NUMBER OF CYCLES. THE NURSE SAID THAT IT DOES APPEAR THAT THE HC SHORTENED THE NUMBER OF CYCLES BUT THAT THE HP'S THERAPY STILL ENDS UP BEING THE SAME, CORRECT EVERY TIME. THE NURSE STATED THEY MADE SURE THE HP WAS COMFORTABLE USING THE HC AND ISN'T JUST "PUSHING BUTTONS" AND THE HP APPEARED PRETTY COMFORTABLE WITH USING THE HC. THE NURSE STATED HP DID NOT SWAPPED OUT THE HC AND DOESN'T BELIEVE HE WANTED TO BECAUSE, HE HAS NEVER REALLY HAD ANY ISSUES WITH IT OVER THE YEARS. THE NURSE STATED THAT THE HP HAS BEEN PERFORMING THERAPY FINE AND HAS HAD NO ADVERSE REACTIONS DUE TO THE HC CHANGING THE CYCLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |